Pb {metRology} | R Documentation |
Lead in wine
Description
A data frame containing reported results for lead (in mg/kg) from CCQM Key Comparison CCQM-K30.
Usage
Pb
Format
A data frame containing 11 reported results with uncertainty data:
- lab
Factor giving abbreviated laboratory identifier
- value
The reported value for lead (mg/kg)
- u
Standard uncertainty (mg/kg). The values in
Pb
were calculated from the reported expanded uncertaintyU
and coverage factork
usingu=U/k
.- k
Coverage factor. Conventionally, the coverage factor is set to a suitable quantile of Student's t based on the Welch-Satterthwaite effective degrees of freedom or simply set to 2 for approximately 95% confidence. In this data set, labs all quoted
k
for approximately 95% confidence.- U
Expanded uncertainty as reported by labs.
- method
Factor indicating general measurement methodology:
- IDMS
Isotope dilution mass spectrometry
- ICP
Inductively coupled plasma spectrometry
- GFAAS
Graphite furnace atomic absorbtion spectrometry
- include
logical; Whether the reported result was included in the calculation of the Key Comparison Reference Value for the study.
Details
The study involved circulation of a homogeneous set of samples of wine for analysis for lead (Pb) content by a number of National Measurement Institutes.
The Key Comparison Reference Value, or assigned value for the lead content, was set at 2.99 mg/kg with expanded uncertainty 0.06 mg/kg.
Source
Hearn, R., Santamaria-Fernandez, R. and Sargent, M. (2008) Final report on key comparison CCQM-K30: Determination of lead in wine. Metrologia 45, 08001, 2008
References
See source.