activityConcentration {oro.pet} | R Documentation |
Calculating SUVs for PET Using QIBA Pseudocode
Description
The standard uptake value (SUV) is calculated based on an 18F-FDG-PET acquistion using ancillary information contained in the DICOM data.
Usage
activityConcentration(pixelData, ...)
## S4 method for signature 'array'
activityConcentration(
pixelData,
CSV = NULL,
seriesNumber = NULL,
method = "qiba"
)
.activityConcentration(
pixelData,
CSV = NULL,
seriesNumber = NULL,
method = "qiba"
)
standardUptakeValue(pixelData, ...)
## S4 method for signature 'array'
standardUptakeValue(
pixelData,
mask = NULL,
CSV = NULL,
seriesNumber = NULL,
method = c("qiba", "user"),
prior = NULL,
decayedDose = NULL
)
.standardUptakeValue(
pixelData,
mask = NULL,
CSV = NULL,
seriesNumber = NULL,
method = c("qiba", "user"),
prior = NULL,
decayedDose = NULL
)
Arguments
pixelData |
is a multidimensional array of signal intensities of class
|
... |
additional arguments |
CSV |
is a |
seriesNumber |
is the SeriesNumber that corresponds to the PET acquisition. |
method |
takes on two possible values ( |
mask |
is a multidimensional array of logical values (only used when
|
prior |
is a list of DICOM header field names that are necessary for
the SUV calculation under |
decayedDose |
is the amount of the RadionuclideTotalDose after being corrected for residual dose in the syringe. This value is NOT usually corrected in the DICOM data. |
Value
A list containing the following items
SUVbwis a multidimensional array, the same dimension as
pixelData
, that contains the standard uptake values.hdris a list of DICOM header fields used in the SUV calculation.
decayTimeis the decay time calculated from the DICOM header information.
decayedDoseis the RadionuclideTotalDose, if taken from the DICOM header information, or the user-specified value.
SUVbwScaleFactoris
\mbox{PatientsWeight}\cdot1000/\mbox{decayedDose}
.
Note
Note, for GE scanners it is common for the RescaleSlope DICOM field to vary on a slice-by-slice basis. This is taken into account if a GE scanner is detected from the Modality DICOM field. However, the InstanceNumber is used to reorder the slices so they match the incoming NIfTI file of PixelData. If this is not correct it may be necessary to manually re-order the RescaleSlope field in the CSV data frame so that the activity concentration is calculated correctly.
Author(s)
Brandon Whitcher bwhitcher@gmail.com
References
https://qibawiki.rsna.org/index.php?title=Standardized_Uptake_Value_(SUV)